COURT OF APPEAL (CIVIL DIVISION)
ON APPEAL FROM THE CHANCERY DIVISION
MR JUSTICE KITCHIN
HC08C00934
[2009] EWHC 1443 (Pat)
Strand, London, WC2A 2LL |
||
B e f o r e :
LORD JUSTICE MOORE-BICK
and
LORD JUSTICE ETHERTON
____________________
COOK BIOTECH INCORPORATED (a company incorporated under the laws of Switzerland) |
Appellant |
|
- and - |
||
EDWARDS LIFESCIENCES AG (a company incorporated under the laws of the State of Indiana, USA) |
Respondent |
____________________
Mr Roger Wyand QC and Mr Piers Acland QC (instructed by Bird & Bird LLP) for the Respondent
Hearing date : 9th June 2010
____________________
Crown Copyright ©
Lord Justice Etherton :
Introduction
The background
"22 … [T]he aortic valve sits in the aortic valve annulus, a fibrous ring at the junction between the left ventricle and the aorta immediately below the sinuses. The aortic valve itself has three leaflets (or cusps) which are half moon shaped. As mentioned, when the left ventricle contracts, the pressure inside the ventricle increases until it is greater than in the aorta, at which point the aortic valve opens. When the ventricular contraction ends, pressure in the left ventricle rapidly drops. When it falls below the pressure in the aorta, the leaflets of the aortic valve collapse and come together along their edges (commissures) and flow of blood from the aorta back to the heart is prevented. Sitting within the aortic sinuses and within a few millimetres of the leaflets are the coronary ostia, which are openings that lead to the coronary arteries. It is crucial that these are not blocked when a valve is replaced because the coronary arteries provide the heart muscle with blood."
"24. Surgeons have been replacing diseased or malfunctioning heart valves for over 40 years. They have used for this purpose a range of prosthetic valves, both mechanical and biological.
25. Mechanical replacement heart valves are generally made from a combination of metal, carbon and plastic and typically provide a valve function through a tilting disc or a ball moving within a cage. They have a long life span but patients suffer an increased risk of thrombus formation which requires them to undergo life-long anticoagulation therapy.
26. Biologically derived (bioprosthetic) valves attempt to replicate more closely the structure and dynamics of a physiological heart valve. They are made of tissue, generally mounted on a textile cuff or metallic or plastic frame and fall into three categories: homograft (human whole valves), xenograft (animal whole valves) and fabricated (valves tailored from animal pericardium, the tissue that covers the outside of the heart). The latter two categories are those of most importance in the context of the present case.
27. Xenograft valves are normally of porcine origin, but can also be of equine or bovine origin. The valve is physically removed from the animal and treated chemically in order to make the biological tissue immunologically inert and sterile and to improve its mechanical properties. It is then attached to a textile cuff allowing it to be sutured to the heart tissue or mounted on a frame which provides some mechanical support. Such a frame is usually referred to as a stent. In 2000, a well established bioprosthetic valve using a porcine valve was called the "Hancock". In this device the valve is fixed to a non-collapsible stent covered with fabric, allowing it to be sutured into the patient around its circumference. The valve has three leaflets which, when the valve is closed, meet at their free margins, that is to say the edges which are free to move from the centre of the valve towards the circumference of the valve when opening and free to return to the centre when closing. The line at which any two of these free edges meet is referred to as the commissure or line of coaption.
28. In the case of fabricated valves, leaflets, normally three in number, are fashioned from a sheet of pericardium and again attached to a frame and sewing cuff. In 2000, one of the most successful fabricated bioprosthetic heart valves was the "Carpentier-Edwards pericardial aortic prosthesis", also known as the "Perimount". It was implanted as early as 1980 in France and approved for use in the US in 1991. It looks like this:
29. Like the Hancock, the Perimount comprises a non-collapsible stent, covered with fabric, which provides a means to suture it into the patient. The stent has three projecting portions known as commissural posts to which the leaflets are connected at the periphery of their commissures. The leaflets are also sutured to the fabric covered stent along the entirety of the inflow side of the valve (the margin of attachment) to preserve the valvular mechanism geometry and ensure the valve does not leak peripherally.
30. Pericardium is the only natural tissue that has ever been used commercially to fabricate a bioprosthetic valve. Its advantages are numerous. It is available in relatively large quantities, permits the production of leaflets that have uniform thickness, strength and flexibility, can be cut to any desired size and, after suitable chemical treatment, possesses physical properties that closely resemble those of the leaflets of human valves. Moreover, pericardium is biocompatible and exhibits low thrombogenicity. For all these reasons, for many years before 2000, pericardium was the only tissue used for the production of commercialised fabricated bioprosthetic valves. However, such valves do suffer from the drawback that they have a tendency to denature or calcify, which affects their long term performance.
31. As Professor Williams [Cook's bioengineering expert] explained, by 2000, attempts had been made to develop a surgically implantable polymer leaflet heart valve which was seen as having the potential to avoid the difficulties of thrombosis caused by the mechanical valves and calcification which tends to occur with bioprosthetic valves. However, no commercial polymer leaflet surgical valves existed at that time."
"32. Surgical heart valve replacement involves a major operation and is not suitable for all patients. However, from the 1960s a new branch of medicine emerged known as interventional cardiology. This is the practice of percutaneously treating problems within the heart and associated vessels and is the province of physicians rather than surgeons. Interventional procedures are carried out using a catheter to access the site in the heart or vasculature where the intervention is to be performed.
33. As Professor Rothman [Cook's expert interventional cardiologist] elaborated, in 1977, Andreas Gruentzig performed the first human balloon angioplasty procedure in which a catheter carrying a balloon was inserted into an occluded human coronary artery and then expanded to force the artery open. By 1990, balloon angioplasty and two related techniques called valvuloplasty (using inflation of a balloon catheter to try to open up a stenotic (narrowed) heart valve and improve blood flow) and atherectomy (using a high speed rotating device or a directional slicing device to remove plaque from the inside of an artery) were regularly being undertaken.
34. Most interventional cardiology is performed percutaneously using a needle inserted into the femoral or radial artery. But it is also possible to cut through the skin over the vessel using a procedure known as "cut-down". Once access to the artery is secured, the catheter is passed to the heart against the blood flow in what is known as a "retrograde" approach. Access to the heart can also be achieved by means of an "antegrade" approach, that is to say passing the catheter in the same direction as the blood flow. In this case the catheter is introduced into a peripheral vein and then advanced along the vena cava to the right side of the heart. If access to the left side of the heart is required then the catheter must be fed through the wall (called the septum) which lies between the two atria of the heart. This technique is used to perform mitral valvuloplasty.
35. In the course of the 1980s and 1990s a great deal of work was also being carried out into the design of expandable stents for translumenal implantation. These were developed to scaffold the internal surface of an artery, initially to prevent an acute closure at the time of a balloon angioplasty procedure, particularly of the coronary artery. However, in the 1990s two major randomised trials known as Benestent and Stress showed that the use of stents also resulted in reduced occurrence of re-narrowing (restenosis) compared with patients receiving balloon angioplasty. As a result, by 2000, stents were being used electively with balloon angioplasty in the majority of cases. They also allowed interventional cardiologists to attempt angioplasty in higher risk and more diseased vessels because they knew that stents had the capacity to prevent short-term and long-term complications.
36. Stents essentially fall into two categories, those which are balloon expandable and those which are self-expanding. Balloon expandable stents are compressed around a balloon and inserted into a peripheral vessel by catheter. Once the balloon expandable stent reaches its destination the balloon is expanded to force open the stent by plastic deformation. The balloon is then deflated and the catheter withdrawn. Some, like the Palmaz-Schatz had a slotted tube design:
37. Others, such as the Gianturco-Roubin had an expandable wire coil design:
38. Self-expanding stents are made of a spring or of a "memory metal" such as nitinol. These require a sheath to maintain the stent in its compressed form during delivery. Once the stent reaches its desired location the sheath is withdrawn and the stent expands. One of the first self-expanding stents was the Gianturco Z-stent, which was first used in the mid-1980s. It has a "zigzag" design and, in a later modification, multiple zigzags were joined together by metal struts or monofilament line to provide a greater degree of stability:
39. By January 2000, many devices consisted of a number of rings joined together and those in the art tended to describe the whole of any such device as a single stent, irrespective of how many rings it might contain.
40. Clearly, stents of different lengths and diameters may be required for different applications. By 2000, it was the general practice to "size" the stent to a diameter approximately 10 to 20% greater than the normal diameter of the treated vessel so as to ensure it would remain in place once deployed and leave the lumen of the vessel unobstructed.
41. Finally, I should mention that for many years prior to 2000 sheaths had been used to cover or contain stents prior to deployment. It was also well known to cover the outside of stents with bio-compatible material such as Dacron, for example to create stent grafts used to support or isolate a weak portion in a vessel, such as an arterial aneurysm."
The Patent
"44. The "Background of the invention" is set out from paragraphs [0003] to [0007]. Two types of known replacement valves are described: mechanical devices with moving ball valves which are susceptible to clot formation and problems associated with long-term wear and tear; and biological valves which suffer from a variety of problems including the supply of valves, immune response and problems associated with positioning. The Patent explains there is therefore a need for alternative and improved devices and methods of providing valvular function within vessels of the body.
45. Paragraph [0008] contains a "Summary of the Invention". It discloses a medical device comprising a frame with a valve located within it. The frame comprises a radially-expandable stent (including especially a self-expanding stent) which can be delivered using a catheter and then deployed and expanded at a target site in a body lumen such as an artery or vein. A preferred use is for the treatment of incompetent veins in the legs or feet.
46. There then follows a "Detailed description of the invention". Paragraph [0010] explains that a valve assembly may have two or more leaflets or cusps. The structure of a typical stent of the invention is depicted in figures 1-3 and described in paragraphs [0011] to [0013]. One embodiment is said to be a self-expanding stent such as the Gianturco and the figures depict a simple arrangement of a cylinder formed by a wire bent or otherwise formed into a zigzag configuration. The specification explains that the bends at the proximal and distal ends of the stent may be connected by sutures which can be used to adjust the size of the stent lumen upon expansion.
47. Paragraphs [0014] to [0017] explain how the valve may be fashioned from a sheet of valve material draped over the stent lumen and then pushed down into its interior. According to the invention, the valve material is said to be a collagen containing bio-material comprising pericardium which is then fixed to the stent frame by a variety of well known means including sutures, adhesives and folds. Connection of the valve to the frame is shown in figures 6A and 6B and the Patent explains that it may be sutured at its distal and proximal ends.
48. Paragraph [0018] relates that once the sutures are generally in place, the valve sheet will form a valve pocket, as shown in figure 6B:
49. The pocket has a valve apex (50) which extends inside the stent lumen and may be sutured to the distal end of the stent frame. There is a part of the valve that will form a central valve portion (49) that is not directly sutured to the stent, but otherwise the valve is sutured around its proximal perimeter to the proximal end of the stent. The valve portion (49) forms the valve opening (52) through which fluid can pass as it flows from the distal to the proximal end of the device. However, if the flow is reversed then the valve pocket (46) fills and the fluid pressure causes the valve portion (49) to extend outwards and, when it does so, to contact the other leaflets or cusps and so form a seal to stop or impede fluid flow.
50. Figure 8, reproduced below, illustrates the valve set in the stent with its distal apex (50) sewn to a distal bend of the stent with a suture (40). It also shows the proximal perimeter of the valve connected to the proximal portion of the stent with two sutures (44):
51. Paragraph [0021] continues that the valve opening is actually created in the final step of preparation of the preferred device. First, a second valve pocket is made by pushing the same sheet of valve material down into the interior on the other side of the stent. The two valve pockets are now sitting side by side. The opening can then be created by cutting a slit in the sheet which can be sized according to the intended flow rate of the passing fluid. The Patent also recognises that opening and closing the valve may cause increased wear and tear at the corners of the opening and, for this reason, reinforcements may be provided in the form of sutures, as illustrated at (53), or by the use of adhesives or any other material or mechanism that permits increased structural integrity. The Patent also explains in paragraph [0033] that the slit may terminate several millimetres (say 1 to 5 mm) before reaching the edge.
52. There then follows a description of how the devices of the invention may be made to different sizes. Paragraph [0022] explains this may be achieved either by elongating the length of the struts of a single stent or by joining a number of stents together (by, for example, sutures). It is preferred that the overall length of the device provides an aspect ratio (length to expanded diameter) sufficiently high to permit proper alignment of the device and that aspect ratios of length to expanded diameter of 1:1 or greater are preferred. It is to be noted, however, that in devices comprising multiple stents there is no requirement that the individual stents should themselves be of any particular length. The teaching of the Patent is simply that the length of the whole device should be appropriate for its intended application and its aspect ratio should be such as to allow proper alignment.
53. A variety of multiple stent structures are then described and depicted but some are said not to be part of the invention, a reflection of the citation by the examiner of the 614 application. Thus figure 12, described in paragraph [0028], is said to be a multi-stent device of the invention:
54. As is figure 17:
….
55. There is one other element of the description to which I should refer. The Patent explains in paragraphs [0029] to [0031] that the outside of the stent may be wholly or partly covered by a sheath. So, for example, excess valve material may be folded over to increase the structural integrity of the device and to present a smoother surface to the body upon implantation. Alternatively the sheath may be made of a synthetic material such as Dacron."
[A] A stent valve, suitable for placement in a vessel, the vessel further having a diameter (84) and an inner lumenal surface, comprising:
[B] a) a radially expandable stent (20) having a proximal stent end (31) and a distal stent end (33),
[C] the stent having an expanded diameter (86) sized to permit contact with an inner lumenal surface of the vessel;
[D] b) a valve (41) having a proximal valve end (48) and a distal valve end (50),
[E] the valve being at least partially located within an inner portion of the stent,
[F] wherein the valve is formed with a collagen containing bio material (38),
characterised in that
[G] said collagen containing bio material comprises pericardium
[H] and extends within said stent (20) substantially from said proximal stent end (31) to said distal stent end (33)
[I] and forms at least two valve leaflets (46)
[J] that extend substantially from said proximal stent end (31) to said distal stent end (33),
[K] with the proximal valve end (48) connected to the proximal stent end (31).
"15. The stent valve of claim 14, wherein the valve opening (52) extends across the diameter of the stent (20) so as to terminate at least 1 mm from a stent perimeter (34)."
"22. The stent valve of claim 21, wherein the valve opening (52) terminates at least 1 mm from a stent perimeter (34)."
Andersen
"113. Andersen is a US patent entitled "Valve prosthesis for implantation in the body and a catheter for implanting such valve prosthesis". It was published on 2 May 1995. As both parties recognise, this publication forms the basis for the strongest obviousness attack on the validity of the Patent.
114. In the section headed "Background of the Invention", it is explained that the invention relates to a valve prosthesis, preferably a cardiac valve prosthesis, for implantation in the body and comprises a collapsible elastical valve mounted on an elastical stent. The "Summary of the Invention" elaborates that the invention provides a stent made from a radially collapsible and re-expandable cylindrical support and a collapsible valve for implantation in the body by means of a catheter. The valve is mounted on the stent by, for example, sutures. The stent itself may be grate shaped, loop shaped or helical and may be either self-expanding or non self-expanding, in which case it may be compressed onto a balloon catheter. The valve prosthesis can be used to replace a natural valve or to establish a new valve function in one of the channels in the body which do not naturally contain a valve. When the valve prosthesis is used as a cardiac valve prosthesis in the aorta, it is explained that it can be mounted in the descending part of the aorta, in a position between the coronary arteries and the left ventricle of the heart, or in the aorta in a position immediately after the mouth of the coronary arteries. It can also be used in the pulmonary artery or the right ventricle for replacing the pulmonary valves.
115. Andersen continues in column 4 that the stent may be made with a relatively great height and with a cylinder surface which is closed by a suitable material. This may facilitate the implantation of the device and securing it in position in the aorta. Such an embodiment is also said to be suitable for a prosthesis inserted into veins which have relatively thin and weaker walls. Here the sheath provides a greater surface over which to distribute the outward pressure necessary to secure it in position.
116. The detailed description of a preferred embodiment begins in column 5 and is illustrated in figures 1 and 2:
117. It can be seen that the stent (1) is made from two surgical stainless steel wires ((2) and (3)) folded into loops. Three loops (4) are higher than the others and intended to secure the commissural points (5) of a biological cardiac valve (6) which is mounted in the stent. The remaining loops form circumferentially expandable sections (25) between the commissural points (5). The two folded wires are bent to form rings ((7) and (8)) which are closed by welding their ends. These are placed on top of each other and are secured by a means of a number of sutures.
118. The description continues that in the particular embodiment described the biological valve (6), a xenograft, was removed from a slaughtered pig, cleaned and then mounted in the stent (1). The valve had an outer diameter of 25-27mm and was mounted in the stent by means of sutures.
119. The whole valve prosthesis was then compressed, so reducing its outer diameter to 10mm and mounted on a balloon catheter with an outer diameter of 13.6mm. As Professor Rothman explained, the size of the valve meant that the prosthesis could only be compressed to a certain extent and, as a result, could not have been percutaneously delivered in a human, where the artery in the groin has a maximum diameter of about 8mm.
120. Figures 5 to 7 show a schematic representation of how the device can be deployed in the aorta by using a catheter (11) and an inflatable balloon (13):
121. Andersen explains at column 6, lines 30 to 36 that to obtain an effective fastening in the aorta, the outer dimension of the prosthesis is greater than the diameter of the aorta. This means that it fits tightly against the inner wall of the aorta with a pressure which is sufficiently large to prevent detachment due to the flow of blood.
122. Various modifications are proposed from column 6, line 66 to column 7, line 16. Specifically it is explained that the prosthesis may be modified and made solely of one closed ring folded in a number of loops or with three or more mutually secured loop-shaped rings placed on top of each other. Moreover it is possible to make the stent having a thread structure which instead of loops is grate shaped, helical or formed in some other way as to permit the compression and expansion of the stent and the fastening of the collapsible valve inside it. It is further explained that instead of a biological valve it might be possible to use other collapsible valves, such as valves made from synthetic materials such as polyurethane, and valves with more or fewer flaps than three.
123. Finally, Anderson reiterates that the stent may have a closed cylindrical surface, making it especially suitable for use in vessels with weak walls, such as veins."
The judgment
(1) (a) Identify the notional "person skilled in the art".
(b) Identify the relevant common general knowledge of that person.
(2) Identify the inventive concept of the claim in question or, if that cannot readily be done, construe it.
(3) Identify what, if any, differences exist between the matter cited as forming part of the "state of the art" and the inventive concept of the claim or the claim as construed.
(4) Ask whether, when viewed without any knowledge of the alleged invention as claimed: do those differences constitute steps which would have been obvious to the person skilled in the art or do they require any degree of invention?
"In a case such as this, involving as it does devices which are not unduly complicated, it is particularly important to assess the question of obviousness without the benefit of hindsight and to keep well in mind that simplicity is no bar to invention. As Laddie J explained in Haberman v Jackel [1999] FSR 683 at 697, the simpler a solution, the easier it is to explain and the easier it is to explain, the more obvious it can appear. This may be unfair to an inventor."
"126. Professor Williams was of the view that the loops should be flexible so as to reduce the forces acting on the valve tissue, in the same way as the commissural posts of the surgically implantable valves such as the Perimount and, moreover, that they might or might not touch the lumen and that that he, as a designer, would make sure they did not. For my part I think it is apparent from figures 5 to 7 that they may well contact the lumen, particularly if the valve is placed into the pulmonary artery or into a vein. However, assuming Professor Williams is right, I have no doubt the skilled person would still regard the loops as part of the stent called for by the Patent for all the following reasons. First, they form an integral part of the upper ring and could not be removed without destroying the integrity of the stent. Second, they support the valve, which is one of the functions of the stent. Third, they would be perceived by the skilled person to be performing a similar role to the commissural posts of the surgically implantable valves, such as the Perimount and, as I have explained, these were part of a frame which was described as a stent. Fourth, and entirely consistently, Andersen itself describes them as forming part of the stent (see, for example, column 5, lines 9 to 28). Fifth, whilst I accept that the loops do not themselves open out as the stent is expanded, they do move radially outwards together with the rest of the stent as the intervening sections (25) open out."
Cook's appeal on obviousness, including claims 15 and 22
Discussion
Claim 1
"The lower ring is circumferentially expandable at least along sections thereof which correspond to the circumferentially expandable sections 25. By using a substantially cylindrical thread structure with projecting apices, a reduction in weight is obtained as compared to a stent which is exclusively with the same loop heights for all the loops."
Claims 15 and 22
"The example is a massive reduction, so I think it is just an example of showing how you might buttress or reinforce the edges. I think that … it throws up the issue that you do need to reinforce the edges in some way otherwise you will end up potentially with an Ionescu-Shiley problem."
Conclusion
Lord Justice Moore-Bick
Lord Justice Jacob